Skip to content

Public FDA record

Z-2562-2021: LYMOL Medical Elite X Class Rigid Bronchoscope System, Part No. BD2410, BB2701, BB2702, BB2401, BB2402, B

Recall record Z-2562-2021 for Lymol Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2562-2021
Company in FDA record
Lymol Medical Corporation
FDA record creation date
Sep 24, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reprocessing instructions were revised November 15, 2020, to reflect new processes and equipment used at consignee user facilities. New instructions were not sent to all consignees of the devices at the time of the revision.

Source facts

Action
On August 9, 2021, the firm began notifying all consignees via email that the reprocessing instructions have changed. Included with each notification is a copy of the Urgent Notice, the revised IFU, and a response form. Customers were advised that the sterilization procedures were updated on November 15, 2020. Customers should discard and replace the obsolete procedures with the revised procedures included with the firm's mailing. The revised instructions have been uploaded to the LYMOL website at https://www.lymolmedical.com/instructions.cfm Customers with questions may contact the firm at Kabrano@lymolmedical.com.
Initiated
2021-07-29
Root cause
Under Investigation by firm
Terminated
2022-10-04
Distribution
US Nationwide distribution.
Recall number
Z-2562-2021
Recall status
Terminated
CDRH record ID
188774
FDA record created
2021-09-24
Recall event number
88485

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗