Open original FDA record ↗Recall record
- FDA identifier
- Z-2563-2014
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Sep 1, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential that test results would be assigned to the wrong Patient ID if cobas IT 1000 receives an ADT ADMIT, TRANSFER or MERGE event message that contains a genuine Patient ID (e.g. 999999), which is the same as what has been configured as the Generic Patient ID (999999 by default).
Source facts
- Action
- Roche sent an Urgent Medical Device Correction letter dated July 31, 2014 via UPS Ground (receipt required) to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to follow the steps provided in the letter to prevent assigning patient results to an incorrect generic patient ID. Please contact the Roche Support Network Customer Support Center at 1-800-440-3638, prompt 5, 24 hours a day, seven days a week if you have questions about the information contained in this UMDC.
- Initiated
- 2014-07-31
- Root cause
- Software design
- Terminated
- 2015-06-16
- Distribution
- US Nationwide Distribution in the states of: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, NE, NH, NJ, NY, OH, OR, PA, PR, SC, TN, TX, VA, VT, WA, WI, and WV.
- Recall number
- Z-2563-2014
- Recall status
- Terminated
- CDRH record ID
- 129001
- FDA classified
- 2014-09-01
- Recall event number
- 68912