Open original FDA record ↗Recall record
- FDA identifier
- Z-2563-2021
- Company in FDA record
- Elekta Inc
- FDA record creation date
- Sep 24, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.
Source facts
- Action
- A Recall notification letter titled, "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" was sent to consignees on 09/14/2021 via email. The letter instructs the consignee of the following: "When using Monaco in offline mode, ensure that the active image set is selected in the workspace control when using the Auto Margin tool." The consignee is to post the notice in a place that is accessible to all users until the action is closed. Furthermore, the consignee is to advise the appropriate personnel, working with this product, on the content of the letter. The firm is requesting that the consignee submit an acknowledgement form as well via the Elekta Care" Community or complete the form and return it to Elekta immediately upon receipt, but no later than within 30 days.
- Initiated
- 2021-09-14
- Root cause
- Software design
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of PA, MI, WI, NY, TN, NJ, IA, TX and the countries of United Kingdom, Turkey, Switzerland, Sweden, Spain, Netherlands, Korea, Thailand, Japan, Italy, Hong Kong, Germany, France, Denmark, China, Canada, Belgium, Bahrain, Australia.
- Recall number
- Z-2563-2021
- Recall status
- Open, Classified
- CDRH record ID
- 189270
- FDA record created
- 2021-09-24
- Recall event number
- 88676