Z-2563-2021: Elekta Monaco - Product Usage: used to make treatment plans for patients with prescriptions for external

Recall record Z-2563-2021 for Elekta Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2563-2021
Company in FDA record
Elekta Inc
FDA record creation date
Sep 24, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.

Source facts

Action
A Recall notification letter titled, "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" was sent to consignees on 09/14/2021 via email. The letter instructs the consignee of the following: "When using Monaco in offline mode, ensure that the active image set is selected in the workspace control when using the Auto Margin tool." The consignee is to post the notice in a place that is accessible to all users until the action is closed. Furthermore, the consignee is to advise the appropriate personnel, working with this product, on the content of the letter. The firm is requesting that the consignee submit an acknowledgement form as well via the Elekta Care" Community or complete the form and return it to Elekta immediately upon receipt, but no later than within 30 days.
Initiated
2021-09-14
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide distribution in the states of PA, MI, WI, NY, TN, NJ, IA, TX and the countries of United Kingdom, Turkey, Switzerland, Sweden, Spain, Netherlands, Korea, Thailand, Japan, Italy, Hong Kong, Germany, France, Denmark, China, Canada, Belgium, Bahrain, Australia.
Recall number
Z-2563-2021
Recall status
Open, Classified
CDRH record ID
189270
FDA record created
2021-09-24
Recall event number
88676

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗