Z-2564-2025: Brand Name: FLEX Monoclonal Mouse Anti-Human CD20cy, Clone L26, Ready-to-Use (Dako Omnis) Model/Catalog N

Recall record Z-2564-2025 for Agilent Technologies Denmark ApS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2564-2025
Company in FDA record
Agilent Technologies Denmark ApS
FDA classification date
Sep 15, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Their is the potential for weak staining which may result in false negative CD20 identification.

Source facts

Action
On August 11, 2025, Agilent Technologies, issued a Urgent Medical Device Correction Notification to affected consignees via E-Mail. Agilent Technologies ask consignees to take the following actions: 1: Please check your inventory and identify if you have the affected product listed in Table 1. Be aware that these products show potential weak staining, which may lead to false negative CD20 identification when used on CLL/SLL tissue samples. 2. If you experience weak staining, please contact Agilent Technical Support for troubleshooting. 3. Please confirm that you have received, read, and understood this Field Safety Notification by completing and signing the enclosed Acknowledgement Form and returning it to fieldactions.notifications@agilent.com. 4. Share this notice with anyone who needs to be aware within your organization and forward to any organization where potentially affected devices may have been transferred.
Initiated
2025-08-11
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution including in the states of Alabama, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Nevada, New Jersey, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Virginia, Washington, West Virginia, Wisconsin.
Recall number
Z-2564-2025
Recall status
Open, Classified
CDRH record ID
215555
FDA classified
2025-09-15
Recall event number
97421

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗