From March 2007 to October 2014, product did not contain updated instructions for use (IFU) to clarify the function of locking mechanism and warnings related to the locking screw.
Source facts
Action
On December 21, 2017 a FIELD SAFETY NOTICE was issued to customers notifying of the corrective action. An attachment was supplied with the letter titled "Minimally Invasive Grower - Instructions For Adjustment that is effective on and after October 14, 2014. Questions or concerns can be directed to your local sales representative.
Initiated
2017-12-21
Root cause
Labeling design
Terminated
2018-10-05
Distribution
NY, FL, KS
Recall number
Z-2567-2018
Recall status
Terminated
CDRH record ID
162447
FDA record created
2018-07-28
Recall event number
79485