Open original FDA record ↗Recall record
- FDA identifier
- Z-2568-2018
- Company in FDA record
- Natus Neurology Inc
- FDA record creation date
- Jul 28, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential to be able to import an ambulatory exam for a patient for which the exam was not started.
Source facts
- Action
- Customers were notified via letter on about 02/01/2018. Instructions to ensure the empty exam is removed appropriately after starting the ambulatory recording included: 1) Only shutdown Recorder via the Recorder UI, either from the File menu or the X button on the top right of the application; 2) If any of the four conditions above occurs and the 2.60 MB exam is not removed, the empty exam can be deleted from NicVue prior to importing the ambulatory exam, doing so will allow the patient ID check to occur normally. Customers were also instructed to provide the recall communication to all applicable users and complete and return the verification form.
- Initiated
- 2018-02-01
- Root cause
- Software design
- Terminated
- 2022-10-21
- Distribution
- Distributed domestically to AR, AZ, CA, CO, DC, FL, GA, ID, IL, IN, MA, ME, MI, MO, N Y, NC, ND, NJ, NV, NY, PA, TN, TX, VA, WA, WI. Distributed internationally to Australia, Belgium, China, Germany, Hong Kong, Israel, Italy, Jordan, Kuwait, Russia, Slovenia, Tunisia, United Arab Emirates, Vietnam.
- Recall number
- Z-2568-2018
- Recall status
- Terminated
- CDRH record ID
- 164231
- FDA record created
- 2018-07-28
- Recall event number
- 80023