Open original FDA record ↗Recall record
- FDA identifier
- Z-2568-2025
- Company in FDA record
- Abiomed, Inc.
- FDA classification date
- Sep 19, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.
Source facts
- Action
- Abiomed notified consignees via letter on about 08/20/2025. Consignees were instructed to quarantine and cease use of all affected units, contact Abiomed Field Service team to initiate the remediation process, complete and return the business reply form with the assistance of your Abiomed Representative, forward the recall notification to any applicable personnel and notify customers if affected units were further distributed or transferred. Abiomed also posted a version of the recall notification letter on their website.
- Initiated
- 2025-08-20
- Root cause
- Device Design
- Distribution
- US distribution to AR, AZ, CA, CO, CT, FL, GA, ID, MA, MI, MO, MS, NJ, NY, OH, PA, TX. International distribution to Canada, Germany, Great Britian, Italy, Netherlands, Norway.
- Recall number
- Z-2568-2025
- Recall status
- Open, Classified
- CDRH record ID
- 215604
- FDA classified
- 2025-09-19
- Recall event number
- 97440