Open original FDA record ↗Recall record
- FDA identifier
- Z-2569-2014
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Sep 22, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential problem with Artis zee and Artis zeego systems running software version VC21B and being used in conjunction with the Large Display, in that under certain circumstances, the release of radiation can become blocked unnecessarily.
Source facts
- Action
- Siemens sent an "Important Customer Safety Notice" dated August 21, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please forward this Safety information to other organizations that could be affected by this measure. If the device has been sold and therefore no longer in your possession, please forward this safety notice to the new owner. We would also request that you inform us of the identity of the device's new owner where possible. Please observe this Customer Safety Notice and comply with the corresponding measures until the update has been fully completed. For further questions please call (610) 219-6300.
- Initiated
- 2014-06-05
- Root cause
- Software design
- Terminated
- 2015-04-21
- Distribution
- Nationwide Distribution
- Recall number
- Z-2569-2014
- Recall status
- Terminated
- CDRH record ID
- 128483
- FDA classified
- 2014-09-22
- Recall event number
- 68704