Skip to content

Public FDA record

Z-2570-2018: XLTEK EMU40EX EEG Headbox

Recall record Z-2570-2018 for Natus Neurology DBA Excel Tech., Ltd. (XLTEK). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2570-2018
Company in FDA record
Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
FDA record creation date
Jul 28, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Possible degradation of an internal electronic component over time which may cause the XLTEK EMU 40EX Breakout box to fail. It may feel hot and plastic case may start melting.

Source facts

Action
Natus Neurology notified customers on or about 05/15/2018, via Advisory Notification Letter to stop using the device immediately and contact Natus regarding replacement options. Customers were advised that if the malfunction occurs, to immediately cease using the device and contact Natus for replacement. Customers were also instructed to communicate the notification to all applicable users and to complete and return the advisory verification form.
Initiated
2018-05-15
Root cause
Other
Terminated
2024-05-06
Distribution
USA Distribution: AL, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, ,RI, SC, TN, TX, UT, VA, WA, WI. World wide Distribution: Canada, China, Cyprus, Germany, Hong Kong, Italy, Mexico, Saudi Arabia, Spain, United Kingdom
Recall number
Z-2570-2018
Recall status
Terminated
CDRH record ID
164948
FDA record created
2018-07-28
Recall event number
80239

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗