Skip to content

Public FDA record

Z-2570-2024: Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. s

Recall record Z-2570-2024 for Merit Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2570-2024
Company in FDA record
Merit Medical Systems, Inc.
FDA classification date
Aug 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

Source facts

Action
On 06/11/2024, the firm sent an "URGENT MEDICAL DEVICE RECALL NOTICE" Letter to customers informing them that specific Merit Medical convenience kits contain plastic syringes manufactured by Jiangsu Shenli Medical Production Co. Ltd. These syringes lack the appropriate/sufficient FDA market authorization. In addition, Merit Medical posted the Urgent Medical Device Recall letter and complete list of affected products/kits on their website at https://www.merit.com/products/documents/product-notices/. Customer are instructed to: 1. Immediately determine if any of the products identified in the attached Customer Response Form (CRF) are within their facility, quarantine them, and discontinue use and distribution. 2. Ensure that appropriate personnel within their organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure this notice is immediately shared with them. Additional distribution details may be required by health authorities. 4. Follow the attached instructions for placing the recall labels on the products containing Jiangsu Shenli syringes, which inform the users to not use the affected syringes. If additional recall labels are needed, contact Customer Service at (801) 208-4381. Product may be returned to inventory once properly over-labeled. 5.At the point of use, the Jiangsu Shenli syringes are to be discarded. Photographs of Jiangsu Shenli syringes are enclosed to assist in identification. 6.Post this URGENT MEDICAL DEVICE RECALL NOTICE and Jiangsu Shenli Syringe photographs on or near the affected products. 7.If replacement syringes are needed, contact Customer Service at (801) 208-4381 for recommended syringe suppliers. For questions contact Merit Sales Representative or Merit Customer Service at 801-208-4381, Hours: 6 am to 6 pm MST Mon.-Fri.
Initiated
2024-06-11
Root cause
No Marketing Application
Distribution
U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Recall number
Z-2570-2024
Recall status
Open, Classified
CDRH record ID
208928
FDA classified
2024-08-12
Recall event number
94797

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗