Recall record Z-2570-2025 for Philips Medical Systems (Cleveland) Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2570-2025
Company in FDA record
Philips Medical Systems (Cleveland) Inc
FDA classification date
Sep 16, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.

Source facts

Action
On 08/05/2025, the firm sent via FedEx an "Urgent Medical Device Correction" Letter informing customers of a software issue affecting the Pinnacle3 Radiation Therapy Planning System performing Advanced Region of Interest (ROI) Expansion and Contraction that results in a potential shift of the intended ROI. The software assumes the patient is always in the Head First Supine (HFS) position when transposing the Advanced Expansion to alternate positions. Customers are instructed to: -When using the Advanced ROI Expansion and Contraction feature in Pinnacle3, only use the Head First Supine (HFS) patient orientation. -Ensure that relevant clinical staff, including physicians, physicists, and dosimetrists, are informed about the potential for directional errors in ROI expansion/contraction under certain orientation conditions. -Customers may continue to use your system(s) in accordance with the intended use and by following the recommendations above. -Circulate this notice to all users of this device so that they are aware of the potential issue. -Retain this letter with their system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. -Complete and return the updated attached response form to Philips promptly and no later than 30 days from receipt of this letter via email to: Philips.recall@philips.com. Note: If customers have returned a response form already in reply to the initial customer letter, complete and sign this updated response form and return to Philips. Philips will provide a label to be affixed to the device that provides instruction to use the HFS position only. A Philips representative will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site to attach /affix the label to the workstation (Reference 2025-EI-TPS-002). Questions-contact your local Philips representative: For North America, contact Philips Customer Care Center at 1-800-722-9377.
Initiated
2025-08-05
Root cause
Software design
Distribution
U.S Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.
Recall number
Z-2570-2025
Recall status
Open, Classified
CDRH record ID
214359
FDA classified
2025-09-16
Recall event number
97049

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗