Open original FDA record ↗Recall record
- FDA identifier
- Z-2571-2025
- Company in FDA record
- BigTree Sales Inc.
- FDA classification date
- Sep 18, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The subject product emits unsafe level of UVC radiation (i.e., the product reaches the exposure limit for Risk Group 3 (high risk) in about 2.2 minutes). The product has a radiation safety defect, in that these radiation emissions are unnecessary to the accomplishment of the product s primary purpose of disinfecting objects and surfaces, and create a risk of skin and/or eye injuries to user and nearby persons.
Source facts
- Action
- BigTree Sales, Inc. notified FDA and their customers as required by 21 CFR 1003.10(b), that their electronic product, Portable UV-C Light Sanitizer Wand/Travel UV Sanitizing Wand, model: HH1103/HD-UV-01PL03 (SKU BT-GERMWAND-LRG-WHT), has been discovered to contain a defect related to its safety. Their customer letter was sent with the knowledge of the U.S. Food and Drug Administration (FDA).
- Initiated
- 2025-08-29
- Root cause
- Radiation Control for Health and Safety Act
- Distribution
- U.S.A
- Recall number
- Z-2571-2025
- Recall status
- Open, Classified
- CDRH record ID
- 216124
- FDA classified
- 2025-09-18
- Recall event number
- 97621