Z-2573-2025: UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520

Recall record Z-2573-2025 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2573-2025
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Sep 17, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.

Source facts

Action
On August 13, 2025, Beckman Coulter initiated a "Urgent Medical Device Recall" notification via E-Mail. Beckman Coulter expanded the recall to additional consignees September 25, 2025 Beckman Coulter asked consignees to take the following actions: For specimens with a WBC count > 30 x 103 and = 90 x 103 cells/L, follow the actions listed within the customer letter. For specimens with a WBC count > 90 x 103 cells/L, follow the actions listed within the customer letter. To identify WBC counts > 30 x 103 cells/L, consider implementing decision rules within your Laboratory Information System (LIS), Middleware, or on affected instruments. Consult your Medical Director to determine if a retrospective review of results is warranted. Beckman Coulter recommends posting this letter on or near the affected analyzers Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of this letter. Acknowledge receipt of the notification.
Initiated
2025-08-13
Root cause
Component change control
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Egypt, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Korea Republic of, Kuwait, Kyrgyzstan, Libya, Lithuania, Mexico, Nepal, Netherlands, New Zealand, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Switzerland, Taiwan, Province of China, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Uruguay, Vietnam.
Recall number
Z-2573-2025
Recall status
Open, Classified
CDRH record ID
215645
FDA classified
2025-09-17
Recall event number
97464

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗