Skip to content

Public FDA record

Z-2573-2026: Surgify Halo, 5.4 mm, Long, Model/Catalog Number: 54.140.SHD.U1; drills, burrs, trephines & accessories (

Recall record Z-2573-2026 for SURGIFY MEDICAL OY. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2573-2026
Company in FDA record
SURGIFY MEDICAL OY
FDA classification date
Jun 24, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS).

Source facts

Action
Surgify issued a Medical Device Correction Notice - Surgify Halo - Restriction of Use notice to its consignees on 4/30/2026 via email. The notice explained the issue, potential risk, and requested the following: "What steps must the user take? Please review your current practice and ensure the device is only used in accordance with the updated safety information. Instruction to Customers: The customers have been instructed to do the following: - Do not use this device in endoscopic surgical procedures. If endoscopic access is required, an alternative instrument should be selected. Specifically: . Do not use this device in endoscopic surgical procedures. If endoscopic access is required, an alternative instrument should be selected. . Ensure that all relevant clinical staff and surgical team members are made aware of this notice. . Please retain this notice for your records and attach it to the relevant product documentation. Please acknowledge receipt of this Medical Device Correction Notice by replying to raqa@surgifymedical.com within 30 business days, confirming that the notice has been read, understood, and communicated to all relevant users within your organization. Please include your facility name, your name, and your role. If you have previously used the device in endoscopic procedures without incident, no patient follow-up is required; however, this application should now be discontinued in line with the updated safety information." For questions regarding this matter, please contact us by phone at +358 105 176 310 or via email at support@surgifymedical.com.
Initiated
2026-04-30
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of MN, MA, IN, LA, CA, NY, FL.
Recall number
Z-2573-2026
Recall status
Open, Classified
CDRH record ID
220197
FDA classified
2026-06-24
Recall event number
98918

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗