Open original FDA record ↗Recall record
- FDA identifier
- Z-2576-2026
- Company in FDA record
- Medtronic Perfusion Systems
- FDA classification date
- Jun 24, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
Source facts
- Action
- An URGENT: MEDICAL DEVICE RECALL notification letter dated 5/12/26 was sent to customers. Customer Actions: Medtronic requests that you take the following actions: " Review your inventory for lot numbers listed in Attachment A. " If you have unused inventory from the listed lot numbers, immediately quarantine and return the products to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return and exchange or credit of this unit. Your Medtronic sales representative can assist you in the return of unit as necessary. " Complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. This form must be returned even if you do not have the affected product in your possession. " Please share this notification with others in your organization as appropriate. If product listed above has been forwarded to another facility, please notify the facility of this Medtronic Urgent Medical Device Recall. " Please maintain a copy of this communication in your records. If you have any questions regarding this communication, please contact Medtronic Field Representative.
- Initiated
- 2026-05-12
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - US Nationwide distribution.
- Recall number
- Z-2576-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220292
- FDA classified
- 2026-06-24
- Recall event number
- 98966