XiO: When the patient is positioned with their feet toward the CT gantry, the coordinates exported to the LAP IsoMark laser positioning system are not correct.
Source facts
Action
The firm, CMS, Inc., sent a "XiO Customer Advisory" letter dated April 8, 2008 to all affected customers. The letter describes the product, the problem, and the actions to be taken ( workaround). The letter also states the problem will be resolved in a future release. No release number or date was given.
If you have any questions, please call 408-380-8023.
Initiated
2008-04-08
Root cause
Software design
Terminated
2014-04-08
Distribution
Worldwide distribution: USA (nationwide) and countries of: China, France, Germany, Japan, Mongolia, Netherlands, Nicaragua, Poland, Russia, South Korea, Switzerland, and Turkey.
Recall number
Z-2577-2011
Recall status
Terminated
CDRH record ID
98677
FDA classified
2011-06-16
Recall event number
58210