Z-2577-2023: BeamAdjust software Version 2.3.3 Ref: S080032 The following devices are affected when a relative calibr

Recall record Z-2577-2023 for PTW-FREIBURG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2577-2023
Company in FDA record
PTW-FREIBURG
FDA classification date
Sep 11, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibration file.

Source facts

Action
PTW North America notified US End users via email and attached Urgent Medical Device Correction letter from PTW Freiberg on 8/23/23. Letter states reason for recall, health risk and action to take: PTW does not recommend performing a relative calibration of the OD 1000 and OD 1600 using Tiff files and DICOM data sets and recommends using alternative reference matrices. A correct result can also be generated if the reference matrix for relative calibration is rotated 90 counterclockwise in VeriSoft and saved before loading into ArrayCal. The information mentioned above should be observed by the customer. No further action is required. Pass this notice on to everyone at your facility/hospital or at another facility/hospital where the equipment is located who needs to know about it.
Initiated
2023-06-07
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-2577-2023
Recall status
Open, Classified
CDRH record ID
202802
FDA classified
2023-09-11
Recall event number
92879

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗