Reason stated in the source
The Paradym CRT 8750 and CRT sonR 8770 device models have a software anomaly, which will cause the device to lose the ability to sense/pace and to deliver further therapy. The software anomaly could only occurs under a specific sequence of events.
Source facts
- Action
- Sorin Group issued a "Medical Device Correction" letter dated June 7, 2010 to consignees. The letter described the product, the problem, and actions to be taken by customers. A Sorin Group representative will contact customers to update the software. Prior to receiving the software update, customers should program the exercise AV delay to 40 ms, which will trigger a reset in the event of an anomaly. This will restore normal operation after shock delivery. If the anolmaly occurs on an implanted unit, a RV coil or SVC coil continuity test should be started, which will ensure that the device recovers proper operation. Corin Group can be contacted at 1-615-788-1807.
- Initiated
- 2010-06-07
- Root cause
- Software design
- Terminated
- 2012-03-29
- Distribution
- Nationwide Distribution in the states of AZ, CA, FL, IA, LA, MA, MS,MO, NJ, NY, NC, OH, OK, PA, SC, TN, TX, and VA.
- Recall number
- Z-2579-2010
- Recall status
- Terminated
- CDRH record ID
- 92939
- FDA classified
- 2010-09-27
- Recall event number
- 56199