Open original FDA record ↗Recall record
- FDA identifier
- Z-2579-2016
- Company in FDA record
- GE Healthcare It
- FDA record creation date
- Aug 17, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
While merging exams in a test instance, merges carry forward in the production system, when the middle tier is configured to the wrong IP address of the Centricity Exam Manager. When a current and historical exam are opened at the same time, the system asynchronously tries to access the operation specifying how each image should be grouped, causing random image-series grouping errors.
Source facts
- Action
- GE sent an Urgent Medical Device Correction letter dated November 15, 2012. The software has been modified to correct these defects. The modified software was released and installed in user sites under GE Healthcare FMI #85196.
- Initiated
- 2012-11-15
- Root cause
- Software design
- Terminated
- 2017-12-13
- Distribution
- Worldwide distribution including US nationwide, Puerto Rico, Australia, Austria, Belgium, Bermuda, CANADA, Denmark, Egypt, ENGLAND, France, Germany, India, Ireland, Hong Kong, Korea, Kuwait, Italy, Malta, Netherlands, Portugal, Qatar, Saudi Arabia, SCOTLAND, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, and UK.
- Recall number
- Z-2579-2016
- Recall status
- Terminated
- CDRH record ID
- 117066
- FDA record created
- 2016-08-17
- Recall event number
- 64785