Open original FDA record ↗Recall record
- FDA identifier
- Z-2579-2026
- Company in FDA record
- Philips North America Llc
- FDA classification date
- Jun 24, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.
Source facts
- Action
- Philips Medical notified consignees on about 06/05/2026 via "URGENT Medical Device Correction" letter. Consignees were notified of the issue and potential hazard and instructed to continue the use of the device but avoid touching the display while the device is booting up and provide this notification to all personnel who need to be aware within each organization or to any organization where affected product(s) have been potentially transferred. Philips will release new software which will correct this problem. A Philips representative will contact you to arrange for an upgrade to the new software.
- Initiated
- 2026-06-05
- Root cause
- Software design
- Distribution
- US Nationwide distribution in the states of AZ, CA, CO, IN, MA, MI, MN, PA, SD, TN.
- Recall number
- Z-2579-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220889
- FDA classified
- 2026-06-24
- Recall event number
- 99126