Open original FDA record ↗Recall record
- FDA identifier
- Z-2582-2023
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Sep 13, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6 transducer and viewing in the Dual format visualization function, which may lead to misdiagnosis of a patient's condition or influence patient management decisions in a negative way.
Source facts
- Action
- On 7/13/23, correction notices were emailed, mailed, or delivered to customers who were asked to do the following: What if I imported my user-generated presets of an 18L6 transducer from an ACUSON Redwood 1.0 system onto an ACUSON Redwood 2.0 system, then used an 18L6 transducer with Dual format to perform patient examinations using my ACUSON Redwood 2.0 system? Review any ultrasound examination results obtained in this situation. Only measurements taken from an 18L6 transducer within Dual format on ACUSON Redwood 2.0 systems are impacted. Ensure that all users of affected ultrasound systems within your organization, and others who may need to be informed, receive the relevant safety information provided with this notice and take the actions specified herein. Your Customer Service Engineer will contact you to schedule a facility visit to update the system or inform you of a remote update when the software update is available. The software update is currently under development and estimated to be available by fall of 2023. Customers with additional questions can contact customer service at 1-800-888-7436.
- Initiated
- 2023-07-13
- Root cause
- Software design
- Distribution
- Worldwide - US Nationwide distribution including in the states of IL, WI, CA, NM, TX, NC, MI, MD, NY, OH, IN, NE, PA, LA, SC, MO, OK, VA, ME, AZ, PR, IA, NH, NJ, AR, FL, SD and the countries of GR, ES, SG, TH, KR, IN, DE, TR, AT, IT, AU, JP, FR, CA, PT, SK, BR, SA, GB, NZ, RO, SE, AE, CN, HR, PL, CH, CR, IQ, NL, MX, ZA, BO, PH, MA, PA, EG, FJ, JO, DZ, TW, DK, CZ, CO, UG, CL, VN, CM, BG, QA, AZ, EC, KW, MY, PE, GG, SV, FO, ID, NP, BA, IL, LT, DO, IE.
- Recall number
- Z-2582-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202424
- FDA classified
- 2023-09-13
- Recall event number
- 92750