Skip to content

Public FDA record

Z-2582-2023: ACUSON Redwood 2.0 ultrasound systems, REF: 11503314

Recall record Z-2582-2023 for Siemens Medical Solutions USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2582-2023
Company in FDA record
Siemens Medical Solutions USA, Inc.
FDA classification date
Sep 13, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6 transducer and viewing in the Dual format visualization function, which may lead to misdiagnosis of a patient's condition or influence patient management decisions in a negative way.

Source facts

Action
On 7/13/23, correction notices were emailed, mailed, or delivered to customers who were asked to do the following: What if I imported my user-generated presets of an 18L6 transducer from an ACUSON Redwood 1.0 system onto an ACUSON Redwood 2.0 system, then used an 18L6 transducer with Dual format to perform patient examinations using my ACUSON Redwood 2.0 system? Review any ultrasound examination results obtained in this situation. Only measurements taken from an 18L6 transducer within Dual format on ACUSON Redwood 2.0 systems are impacted. Ensure that all users of affected ultrasound systems within your organization, and others who may need to be informed, receive the relevant safety information provided with this notice and take the actions specified herein. Your Customer Service Engineer will contact you to schedule a facility visit to update the system or inform you of a remote update when the software update is available. The software update is currently under development and estimated to be available by fall of 2023. Customers with additional questions can contact customer service at 1-800-888-7436.
Initiated
2023-07-13
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution including in the states of IL, WI, CA, NM, TX, NC, MI, MD, NY, OH, IN, NE, PA, LA, SC, MO, OK, VA, ME, AZ, PR, IA, NH, NJ, AR, FL, SD and the countries of GR, ES, SG, TH, KR, IN, DE, TR, AT, IT, AU, JP, FR, CA, PT, SK, BR, SA, GB, NZ, RO, SE, AE, CN, HR, PL, CH, CR, IQ, NL, MX, ZA, BO, PH, MA, PA, EG, FJ, JO, DZ, TW, DK, CZ, CO, UG, CL, VN, CM, BG, QA, AZ, EC, KW, MY, PE, GG, SV, FO, ID, NP, BA, IL, LT, DO, IE.
Recall number
Z-2582-2023
Recall status
Open, Classified
CDRH record ID
202424
FDA classified
2023-09-13
Recall event number
92750

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.