Open original FDA record ↗Recall record
- FDA identifier
- Z-2583-2010
- Company in FDA record
- Synthes USA (HQ), Inc.
- FDA record creation date
- Jun 26, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A potential for the implant holder tip to break off and lodge in the SynFix-LR implant, which pose a potential risk to the patient.
Source facts
- Action
- Synthes Spine Hospital sent an ' URGENT: MEDICAL DEVICE LABELING CORRECTION" letter dated August 20, 2010 to all customers. The letter described the labeling issue and actions to be taken by the customers. The customers were instructed to review the supplementary labeling provided; contact their Synthes Spine Sales Consultant to arrange for provision of the Holding Forceps (p/n 388.401) prior to each SynFix-LR procedure; and complete and return the enclosed Verification Section form to Synthes via fax at 610-251-9005 or email: FieldAction@Synthes.com. If you have any questions, please call 1-800-620-7025 x 3500 or contact your Synthes Spine Sales Consultant.
- Initiated
- 2010-07-23
- Root cause
- Error in labeling
- Terminated
- 2014-12-18
- Distribution
- Worldwide Distribution: USA and the countries of Canada, Japan, Switzerland, Australia, Mexico, Malaysia, New Zealand, and Singapore.
- Recall number
- Z-2583-2010
- Recall status
- Terminated
- CDRH record ID
- 93275
- FDA record created
- 2015-06-26
- Recall event number
- 56343