Skip to content

Public FDA record

Z-2584-2010: WECK IMA/ENT Blade Electrode (1.9 cm), REF 809335, Rx Only Sterile, Teleflex Medical, RTP, NC The blade

Recall record Z-2584-2010 for Telefelx Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2584-2010
Company in FDA record
Telefelx Medical
FDA classification date
Sep 27, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Complaints were received regarding peeling and melting of the insulation of the cautery tip of the IMA/ENT blade electrodes.

Source facts

Action
Teleflex Medical sent an "Urgent Medical Device Recall" letter dated July 30, 2010, to all customers. The letter described the product, problem and actions to be taken by customers. The customers were instructed to immediately discontinue use, quarantine any affected products and return all of the affected product to Teleflex Medical; forward the letter to their customers to retrieve product from those locations and complete and return the Recall Acknowledgement Form via fax at 1-866-804-9881. A second mailing to non-responding customers on the consignee list will be mailed approximately four to six seeks after the first mailing. A third mailing to non-responding customers on the consignee list will be made approximately four to six weeks after the second mailing or as an alternative, they may be contacted by phone, e-mail, or fax. Note: A record of these contacts will be maintained. Affected products received by Teleflex Medical will be destroyed upon receipt. Records will be maintained accordingly. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990.
Initiated
2010-07-30
Root cause
Other
Terminated
2011-04-21
Distribution
Worldwide distribution: USA and countries of Canada, Australia and Singapore.
Recall number
Z-2584-2010
Recall status
Terminated
CDRH record ID
93337
FDA classified
2010-09-27
Recall event number
56379

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗