Skip to content

Public FDA record

Z-2585-2025: i-STAT EG6+ cartridge. List Number: 03P77-25.

Recall record Z-2585-2025 for Abbott Point Of Care Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2585-2025
Company in FDA record
Abbott Point Of Care Inc.
FDA classification date
Sep 17, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.

Source facts

Action
Abbott notified consignees via email and letter on about 08/21/2025. Consignees were instructed to download, review and implement the updated i-STAT CG8+, EG7+ and EG6+ cartridge Instructions for Use (IFU), discontinue use of i-STAT CG8+, EG7+ and EG6+ cartridges for testing of capillary whole blood samples for potassium and ionized calcium, and use venous or arterial whole blood or an alternate method for testing capillary potassium and ionized calcium. Update the pH and PO2 reportable ranges on the i-STAT 1 analyzers running i-STAT CG8+, EG7+ or EG6+ cartridges. As this feature is not customizable through the i STAT 1 handheld keypad, please see the i STAT DE Quick Reference Guide - Customizing Reportable Ranges (ART 770547-00 Revision A) for more information. To use the Custom Reportable Range feature, you must have the i-STAT 1 handheld and i STAT DE version 2.3 or higher. Changes to the reportable range may require a change to the Laboratory Information System (LIS) interface. Additionally, consignees were requested to confirm receipt and understanding of this communication by responding to the business reply card included with the letter. If consignees have forwarded any i-STAT CG8+, EG7+ or EG6+ cartridges to another facility, they were requested to notify them.
Initiated
2025-08-21
Root cause
No Marketing Application
Distribution
US Nationwide distribution.
Recall number
Z-2585-2025
Recall status
Open, Classified
CDRH record ID
215609
FDA classified
2025-09-17
Recall event number
97443

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗