Z-2586-2010: Stryker Epic II Critical Care Bed, Model 2030/2031, Stryker Medical, Kalamazoo, MI. Critical Care Patient

Recall record Z-2586-2010 for Stryker Medical Div. of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2586-2010
Company in FDA record
Stryker Medical Div. of Stryker Corporation
FDA classification date
Sep 28, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The Fowler mechanism may malfunction, preventing it from being lowered into a flat position.

Source facts

Action
Stryker Medical sent an Urgent Medical Device Alert letter dated May 14, 2010 to their customers. The letter identified the product, the problem, and the actions customers should take. The letter instructed users to do the following: " Return the postcard included with the letter to confirm receipt of the notification. " Locate the recalled beds, verify the Fowler mechanism is functioning properly and does not display uneven motion when raising and lowering, and is not unusually noisy. " If the Fowler mechanism is not operating correctly, they were to call Stryker at 1-800-STRYKER, option 1, Reference RA-2010-014 to obtain service for the bed. " If any beds have been loaned or sold, the customer is to forward a copy of the notice to the new users and advise the firm of the new location. " Provide the firm with the serial numbers for any beds that were disposed of. " Retain the revised maintenance manual included with the notification. If the customer had any urgent questions or concerns, they were instructed to call the firm at 800-STRYKER, option 1, reference RA-2010-014, Monday-Friday 8am to 5pm (EST).
Initiated
2010-05-14
Root cause
Component change control
Terminated
2011-06-13
Distribution
Worldwide Distribution - USA (all states), and including the countries of ARGENTINA, BRAZIL, CHILE, CHINA, CANADA, MEXICO, GREECE, JAPAN, KOREA, PUERTO RICO, SPAIN, POLAND, TAIWAN, SINGAPORE, INDIA, SWITZERLAND, UK and MALAYSIA.
Recall number
Z-2586-2010
Recall status
Terminated
CDRH record ID
93824
FDA classified
2010-09-28
Recall event number
56507

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗