Open original FDA record ↗Recall record
- FDA identifier
- Z-2586-2024
- Company in FDA record
- EYE COMFORT CARE LLC
- FDA classification date
- Aug 14, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a change in the safety clips for the heated eye masks, the wires to the safety clips may detach as a result of tugging and pulling and may cause a short in the heating element of the mask and could potentially cause burn injuries.
Source facts
- Action
- Between 02/08/2024 to 05/18/2024, the firm send via email, a Stop Sale Notice and an "Voluntary Product Recall Notice" for it UNCLOG Heated Eye Masks (Lot Model# AG23JN24) inform customers that an unauthorized material substitution was made in the manufacturing process of these masks. Additionally, it was identified that a smaller safety clip was used and could compromise the integrity of the mask. Customers are instructed to: 1. Cease Use Immediately. Stop using the UNCLOG Heated Eye Mask to avoid any potential risks. 2. Claim a replacement. The firm is committed to replacing their current masks with a new, fully tested new UNCLOG eye mask at no additional cost. Any questions or assistance, contact Vivian Surya, Founder/CEO of Eye Comfort Care LLC at 888-781-2811 or email at vivian@eyeconfortcare.com
- Initiated
- 2024-02-08
- Root cause
- Vendor change control
- Distribution
- U.S. Nationwide distribution in the states of AL, AZ, CA, CT, FL, KS, KY, MA, ME, MT, NC, ND, NJ, NM, NY, OH, OR, PA, SC, TX, VA, WA, and WI.
- Recall number
- Z-2586-2024
- Recall status
- Completed
- CDRH record ID
- 208839
- FDA classified
- 2024-08-14
- Recall event number
- 94929