Skip to content

Public FDA record

Z-2586-2024: Brand Name: UNclog Eye Mask Product Name: UNclog Eye Mask Model/Catalog Number: AG23JW24 Serial # ag22 Pr

Recall record Z-2586-2024 for EYE COMFORT CARE LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2586-2024
Company in FDA record
EYE COMFORT CARE LLC
FDA classification date
Aug 14, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a change in the safety clips for the heated eye masks, the wires to the safety clips may detach as a result of tugging and pulling and may cause a short in the heating element of the mask and could potentially cause burn injuries.

Source facts

Action
Between 02/08/2024 to 05/18/2024, the firm send via email, a Stop Sale Notice and an "Voluntary Product Recall Notice" for it UNCLOG Heated Eye Masks (Lot Model# AG23JN24) inform customers that an unauthorized material substitution was made in the manufacturing process of these masks. Additionally, it was identified that a smaller safety clip was used and could compromise the integrity of the mask. Customers are instructed to: 1. Cease Use Immediately. Stop using the UNCLOG Heated Eye Mask to avoid any potential risks. 2. Claim a replacement. The firm is committed to replacing their current masks with a new, fully tested new UNCLOG eye mask at no additional cost. Any questions or assistance, contact Vivian Surya, Founder/CEO of Eye Comfort Care LLC at 888-781-2811 or email at vivian@eyeconfortcare.com
Initiated
2024-02-08
Root cause
Vendor change control
Distribution
U.S. Nationwide distribution in the states of AL, AZ, CA, CT, FL, KS, KY, MA, ME, MT, NC, ND, NJ, NM, NY, OH, OR, PA, SC, TX, VA, WA, and WI.
Recall number
Z-2586-2024
Recall status
Completed
CDRH record ID
208839
FDA classified
2024-08-14
Recall event number
94929

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗