Nidek Inc. received information from our manufacturer NIDEK CO. Japan that the shown axial
lengths were much shorter at AL-Scan measurement.
Source facts
Action
Dear Doctor/Recall letters were sent to affected customers on July 7, 2016. The letters describe the issue and provide instruction should the issue occur. The letters inform users that Nidek will reach out in the next few weeks to schedule installation of new software that will solve the problem.
Questions or concerns may be directed to:
Customer Service 1.800.722.0219; Monday- Friday 8 am to 5pm PST
E-mail: AL-Scan_support@noritsuservice.com.
Initiated
2016-07-01
Root cause
Software in the Use Environment
Terminated
2016-12-22
Distribution
US Nationwide Distribution in the states of: NY, PA. GA. AZ, VA, DE, ME,. MO, TN, and NM.
Recall number
Z-2587-2016
Recall status
Terminated
CDRH record ID
148011
FDA record created
2016-08-18
Recall event number
74595