A damaged foot switch may cause instances where no or only intermittent x-ray radiation initiation is possible.
Source facts
Action
US customers were notified via certified mail via USPS. OUS customers were notified through the Philips Markets Organizations.
Philips is issuing two Urgent Medical Device Correction letters, one for Allura and Azurion customers and another one for MultiDiagnost Eleva Customers. Additionally, an IFU addendum will be provided to customers that includes instructions on how to handle the footswitch cable. For Allura and Azurion customers an updated preventative maintenance manual update is also being provided.
Philips will contact customers to schedule a visit to inspect the foot switch and to provide an IFU Addendum.
Initiated
2023-08-15
Root cause
Device Design
Distribution
Worldwide distribution
Recall number
Z-2587-2023
Recall status
Open, Classified
CDRH record ID
202848
FDA classified
2023-09-14
Recall event number
92901