Z-2588-2017: Monica Novii Wireless Patch System product Usage: Novii Interface - The Novii Interface is an accessory

Recall record Z-2588-2017 for MONICA HEALTHCARE LTD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2588-2017
Company in FDA record
MONICA HEALTHCARE LTD
FDA classification date
May 25, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Monica Novii Wireless Patch System-Potential battery damage and excess heat through prolonged charging.

Source facts

Action
The recall strategy outlined below addresses the depth of the recall, plan for public notification, and recall notification (customer letter) effectiveness check. This issue will be resolved by sending an Urgent Medical Device Correction customer letter (attached) to customers with affected units. This letter will inform the customer of the issue and provide safety instructions for immediate mitigation of the issue. The long-term solution is to correct all affected Products at no cost to the customer. The plan for public notification is to contact all customers with affected systems by means of an Urgent Medical Device Correction via e-mail. This letter will be sent to the following titles within the affected accounts: Nurse Managers, Labor & Delivery/Bio-medical Engineering Department Managers/Risk Management Directors. We will track all service requests for the field action to completion. These service requests will record the repair of the affected units. Exceptions to completion of the service request will be recorded as well. Records of field action completion will be retained per our processes and procedures
Initiated
2017-04-28
Root cause
Nonconforming Material/Component
Terminated
2018-03-01
Distribution
The product was distributed thru the United States. Please see attached excel sheet for detailed information about the consignees.
Recall number
Z-2588-2017
Recall status
Terminated
CDRH record ID
155680
FDA classified
2017-05-25
Recall event number
77313

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗