Open original FDA record ↗Recall record
- FDA identifier
- Z-2588-2021
- Company in FDA record
- ARJOHUNTLEIGH POLSKA Sp. z.o.o.
- FDA record creation date
- Sep 29, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The IndiGo power cord might wear during use which could lead to its damage, resulting in the cord being cut, burnt or short circuit with limited emission of sparks.
Source facts
- Action
- The Field Safety Notice dated 6/9/2021 was issued in two different mailings, one on 6/10/2021 and the second on 6/18/2021. The letter explained the issue and provided instructions for use until the product is corrected by Arjo personnel. All caregivers and users of the bed with IndiGo were to be made aware of the Field Safety Notice. A Customer Response Form was enclosed for completion to indicate the number of beds currently listed at the facility and that the customer has read and understands the Field Safety Notice. If the bed has been sold or moved, the customer is to provide information on the new facility.
- Initiated
- 2021-06-10
- Root cause
- Nonconforming Material/Component
- Terminated
- 2026-02-03
- Distribution
- US Nationwide distribution in the states of CA, FL, IL, KY, MD, NV, NY, PA, and SD.
- Recall number
- Z-2588-2021
- Recall status
- Terminated
- CDRH record ID
- 188736
- FDA record created
- 2021-09-29
- Recall event number
- 88469