Skip to content

Public FDA record

Z-2588-2021: Arjo Citadel beds originally assembled with the IndiGo modules during manufcturing, model number CX812A3J

Recall record Z-2588-2021 for ARJOHUNTLEIGH POLSKA Sp. z.o.o.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2588-2021
Company in FDA record
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
FDA record creation date
Sep 29, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The IndiGo power cord might wear during use which could lead to its damage, resulting in the cord being cut, burnt or short circuit with limited emission of sparks.

Source facts

Action
The Field Safety Notice dated 6/9/2021 was issued in two different mailings, one on 6/10/2021 and the second on 6/18/2021. The letter explained the issue and provided instructions for use until the product is corrected by Arjo personnel. All caregivers and users of the bed with IndiGo were to be made aware of the Field Safety Notice. A Customer Response Form was enclosed for completion to indicate the number of beds currently listed at the facility and that the customer has read and understands the Field Safety Notice. If the bed has been sold or moved, the customer is to provide information on the new facility.
Initiated
2021-06-10
Root cause
Nonconforming Material/Component
Terminated
2026-02-03
Distribution
US Nationwide distribution in the states of CA, FL, IL, KY, MD, NV, NY, PA, and SD.
Recall number
Z-2588-2021
Recall status
Terminated
CDRH record ID
188736
FDA record created
2021-09-29
Recall event number
88469

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗