Open original FDA record ↗Recall record
- FDA identifier
- Z-2588-2023
- Company in FDA record
- BioFire Diagnostics, LLC
- FDA classification date
- Sep 30, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel
Source facts
- Action
- On 4/25/23 correction notices were mailed to customers who were told to update the GI pouch module software. Instructions for installation can be downloaded here: https://www.biofiredx.com/e-labeling/ITIFA20GI21. This technical note and download cannot be opened via Firefox browser; please use a different browser. Once the pouch module software has been updated, complete and return the webform located at: https://www.biofiredx.com/gisoftware-5747 or return the acknowledgement form via email to FCA5747@biomerieux.com Questions or concerns can be directed to Technical Support at biofiresupport@biomerieux.com or via telephone by dialing 1.800.735.6544 and selecting option 5.
- Initiated
- 2023-04-25
- Root cause
- Software design
- Distribution
- US:DE, AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: AD, AE, AI, AR, AT, AU, BD, BE, BF, BG, BH, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EE, ES, FI, FR, GA, GB, GE, GH, GM, GR, GU, HK, HU, IC, ID, IE, IL, IN, IQ, IT, JP, KE, KR, KW, LV, MO, MX, MY, NL, NO, NP, NZ, PG, PH, PL, PT, QA, RS, SA, SE, SG, SI, TH, TN, TR, TW, VN, ZA, ZW
- Recall number
- Z-2588-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202812
- FDA classified
- 2024-09-30
- Recall event number
- 92885