Skip to content

Public FDA record

Z-2588-2023: FILMARRAY Gastrointestinal (GI) Panel (Part No.: RFIT-ASY-0104 & RFIT-ASY-0116) Pouch Module Software

Recall record Z-2588-2023 for BioFire Diagnostics, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2588-2023
Company in FDA record
BioFire Diagnostics, LLC
FDA classification date
Sep 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel

Source facts

Action
On 4/25/23 correction notices were mailed to customers who were told to update the GI pouch module software. Instructions for installation can be downloaded here: https://www.biofiredx.com/e-labeling/ITIFA20GI21. This technical note and download cannot be opened via Firefox browser; please use a different browser. Once the pouch module software has been updated, complete and return the webform located at: https://www.biofiredx.com/gisoftware-5747 or return the acknowledgement form via email to FCA5747@biomerieux.com Questions or concerns can be directed to Technical Support at biofiresupport@biomerieux.com or via telephone by dialing 1.800.735.6544 and selecting option 5.
Initiated
2023-04-25
Root cause
Software design
Distribution
US:DE, AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: AD, AE, AI, AR, AT, AU, BD, BE, BF, BG, BH, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EE, ES, FI, FR, GA, GB, GE, GH, GM, GR, GU, HK, HU, IC, ID, IE, IL, IN, IQ, IT, JP, KE, KR, KW, LV, MO, MX, MY, NL, NO, NP, NZ, PG, PH, PL, PT, QA, RS, SA, SE, SG, SI, TH, TN, TR, TW, VN, ZA, ZW
Recall number
Z-2588-2023
Recall status
Open, Classified
CDRH record ID
202812
FDA classified
2024-09-30
Recall event number
92885

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗