Open original FDA record ↗Recall record
- FDA identifier
- Z-2589-2018
- Company in FDA record
- Spacelabs Healthcare, Ltd.
- FDA classification date
- Aug 10, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Arkon Anesthesia Delivery System may go into a failed state (mechanical ventilation ceases) while the machine is in use or while idle.
Source facts
- Action
- The firm, Spacelabs Healthcare, sent "Field Corrective Action Display Unit Resets" letters dated 7/3/2018 and 7/9/2018 to its customers via priority service on 07/11/2018. The letters informed them to circulate this notification to all affected persons and to add copies to appropriate user manuals. The firm will contact customers to replace the display assembly.. Customers with questions are encouraged to call Technical Support at 1-425-363-5212.
- Initiated
- 2018-07-11
- Root cause
- Nonconforming Material/Component
- Terminated
- 2024-03-20
- Distribution
- Worldwide Distribution: U.S (nationwide) to states of.: LA, NE, CO, WY, MT, NM, IN, ME, NY, IL, MA, TX, PA, PR; and countries (OUS) of: PERU BOLIVIA, PANAMA, MALAYSIA, POLAND, IRELAND, MOROCCO, MEXICO, CANADA, UNITED KINGDOM, FRANCE, and CHINA, ECUADOR.
- Recall number
- Z-2589-2018
- Recall status
- Terminated
- CDRH record ID
- 165550
- FDA classified
- 2018-08-10
- Recall event number
- 80462