Open original FDA record ↗Recall record
- FDA identifier
- Z-2589-2021
- Company in FDA record
- Atrium Medical Corporation
- FDA record creation date
- Sep 29, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The product package may potentially contain a mismatched graft that does not contain the radial support ring (helix) identified in the labeling.
Source facts
- Action
- On August 25, 2021, the firm issued Urgent Medical Device - Removal letters to affected customers. Customers were instructed to do the following: - Please examine your inventory immediately to determine if you have any of the Atrium Advanta VXT Vascular Grafts with Part Number/REF Number 22059. - If you have affected product with Part Number/REF Number 22059, please remove the affected products from areas of use and return the product(s) to Getinge following the instructions in the letter. - Should you have affected product you are eligible for either a replacement or credit. - Please forward this information to all current and potential Atrium Advanta VXT Vascular Graft users within your hospital / facility. - If you are a distributor who has shipped any affected products to customers, please forward the recall notification to their attention for appropriate action. - Please contact Getinge Customer Service at (603) 880 1433 to request a return authorization (RMA) and shipping instructions to return any affected product. - Pack the product to be returned with the appropriate return documents and, using the shipping instructions provided by Getinge Customer Service, arrange for pickup with the designated delivery service provider. If you have any questions, please contact your Atrium/Getinge representative or call the Atrium/Getinge Customer Support at (603) 880 1433, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
- Initiated
- 2021-08-25
- Root cause
- Under Investigation by firm
- Terminated
- 2024-04-10
- Distribution
- Worldwide distribution - US Nationwide distribution in the state of Florida and the countries of South Korea and Poland.
- Recall number
- Z-2589-2021
- Recall status
- Terminated
- CDRH record ID
- 189079
- FDA record created
- 2021-09-29
- Recall event number
- 88606