Open original FDA record ↗Recall record
- FDA identifier
- Z-2591-2017
- Company in FDA record
- Merge Healthcare, Inc.
- FDA record creation date
- Jun 15, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.
Source facts
- Action
- Merge sent an Urgent Medical Device Recall letter dated May 5, 2017, to all affected customers. Customers were instructed to be aware of the following workaround: Do not have the simple close study red x icon available in the reading physicians toolbars. This will force them to set status on close which also presents the warning Not all images have been reviewed, do you want to set status anyway and close study or continue reviewing unviewed images? YOUR RESPONSE TO THIS NOTIFICATION IS REQUIRED By replying to this notification, you will be taking the necessary steps to ensure that the issue listed in this document will be resolved. Please reply using the enclosed form and the return addressed envelope. Customers with questions were instructed to send an email to recall@merge.com. For questions regarding this recall call 877-741-6369.
- Initiated
- 2017-05-08
- Root cause
- Software design (manufacturing process)
- Terminated
- 2019-08-08
- Distribution
- Nationwide Distribution
- Recall number
- Z-2591-2017
- Recall status
- Terminated
- CDRH record ID
- 155552
- FDA record created
- 2017-06-15
- Recall event number
- 77249