Open original FDA record ↗Recall record
- FDA identifier
- Z-2591-2018
- Company in FDA record
- ELITech Clinical Systems SAS
- FDA record creation date
- Aug 1, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
ELITech Clinical Systems SAS initiated this recall because some users of ENVOY 500 ISE CALIBRATOR Kit (Part number: 55117), for Envoy 500 systems are observing trouble while calibrating with some vials. Users contacted the firm when they received the messages "ISE OUT OF REPRODUCIBILITY" or "ISE SLOPE OUT OF RANGE" (i.e., failed calibration). This has resulted in delay in obtaining patient results until the calibrator lot is replaced.
Source facts
- Action
- ELITech Clinical Systems SAS notified consignees on April 18, 2018 via USPS certified letter, dated April 10, 2018 titled Product Recall Notice. The Product Recall Notice provided customers/consignees reason for recall, actions to be taken, and included a confirmation of receipt of product recall notice form to be filled out and returned by customer/consignee.
- Initiated
- 2017-04-18
- Root cause
- Under Investigation by firm
- Terminated
- 2020-04-17
- Distribution
- U.S. only. No gov. accounts. No international.
- Recall number
- Z-2591-2018
- Recall status
- Terminated
- CDRH record ID
- 165198
- FDA record created
- 2018-08-01
- Recall event number
- 80321