Z-2592-2017: Proteus 235

Recall record Z-2592-2017 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2592-2017
Company in FDA record
Ion Beam Applications S.A.
FDA classification date
Jun 19, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It has been observed during testing that there could be clinically non-negligible amount of dark current (a well-known physics effect observed in cyclotron) in Pencil Beam Scanning (PBS) treatment mode when system is not irradiating between layers.

Source facts

Action
Ion Beam Application S.A. (IBA) will bring the defect into compliance: 1. Providing recommendations for user actions which, if followed, will prevent dark current. In order to prevent the error from occurring, the IBA customers are provided with a specific QA map that would detect dark current and instructions on how and when to use it (Annex 1). This guidance was included in IBAs User Notice which was sent to users on February 9, 2017. 2. IBA is developing a long term solution consisting of a new IC-cyclo (ionization chamber located at the cyclotron exit) and a Beam Absence Electronic Unit (BAEU) along with a software upgrade. This solution will detect presence of dark current between irradiations. It will prevent starting a new irradiation if dark current is present. This solution is currently in test phase and will be completed in two years. 3. Manufacturer will provide the FDA with progress reports on the effectiveness of the plan, including the number of electronic products repaired on a quarterly basis. The first progress report will be provided to the FDA by September 15, 2017 to the correspondence identified below. CDRH approves the CAP subject to the following conditions: 1.Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2.Any future customer notification letters to be sent must comply with the requirements of 21 CFR 1003.21. This includes, but is not limited to, the prescribed packaging and the statement that you will remedy the defect or bring the product into compliance at no charge. For further questions please call (571) 449-4992.
Initiated
2017-02-09
Root cause
Device Design
Terminated
2023-07-14
Distribution
Worldwide Distribution - US Distribution to the states of : MA, FL, VA, PA, IL, NJ, WA, TN, LA and TX., and to the countries of: Germany, South Korea, France, Czech Republic, Italy, Poland and Sweden.
Recall number
Z-2592-2017
Recall status
Terminated
CDRH record ID
155167
FDA classified
2017-06-19
Recall event number
77175

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗