Skip to content

Public FDA record

Z-2592-2021: Atrium Pneumostat Chest Drain Valve, Part Number 16100

Recall record Z-2592-2021 for Atrium Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2592-2021
Company in FDA record
Atrium Medical Corporation
FDA record creation date
Sep 29, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm is revising the Indication for Use, Contraindications, Warnings, and Precautions, after the device was involved in the treatment of a preterm infant who later died.

Source facts

Action
On August 26, 2021, the firm issued Urgent Medical Device Correction letters to affected customers. Customers were informed that Atrium/Getinge is in the process of updating the Instructions for Use. In the meantime, the following information was provided: REVISED Indication for Use: Pneumostat is indicated for use with adult patients in a variety of medical conditions in which evacuation of air with or without ancillary fluid is required from the pleural cavity and is also designed to facilitate ambulation. These medical conditions may include pneumothorax, thoracic injury, and other related conditions requiring air evacuation. NEW Contraindication: Not for use in pediatric patients. NEW section for Target Population: The Pneumostat is intended for adult patients who require gravity drainage. NEW Warning: Do not use in situations that require active negative pressure in the pleural space. NEW Precaution: The use of the Pneumostat is restricted to the clinical setting. NEW Precaution: The use of the Pneumostat is restricted to professional healthcare users familiar with thoracic surgical procedures and techniques, including the use of thoracic chest drains. Actions to be taken by Customer: - Please examine your inventory immediately to determine if you have any of the Atrium Pneumostat Chest Drain Valves with the product code 16100. - Please ensure that all Pneumostat Chest Drain Valve users at your facility are aware of this notice and post a copy of the Warning Label included with the letter in all inventory locations within your facility where the Atrium Pneumostat Chest Drain Valves are stored. If you have any questions, please contact your Atrium/Getinge representative or call the Atrium/Getinge Customer Support at (603) 880-1433 (press option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Initiated
2021-08-26
Root cause
Under Investigation by firm
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Recall number
Z-2592-2021
Recall status
Open, Classified
CDRH record ID
188689
FDA record created
2021-09-29
Recall event number
88446

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗