Z-2592-2023: Membrane Changer Single-Use, REF MC, each package contains 3 changers; Membrane Changer Reloadable, REF M

Recall record Z-2592-2023 for SenTec AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2592-2023
Company in FDA record
SenTec AG
FDA classification date
Sep 15, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During rework of components during the manufacturing process, an additional piece of material may have been left in the device, which may result in an impact on measurement performance.

Source facts

Action
The recalling firm notified Sentec distributors in a letter flagged Guidance for Sentec Distributors dated 7/25/2023 via email on 7/25/2023 (letter #RF-013993-a). The letter listed the devices impacted, the defect, how it manifests itself, and the risks (impact on measurement readings, impact on the sensor membrane, and harm to patient skin). Actions for the distributor and subdistributor level were provided. Subdistributors were to be informed who received the affected products by forwarding them this guidance document and related attachments using a read receipt on the email. The distributor is to follow-up with each subdistributor to ensure they understand the information provided and that each subdistributor will implement all actions as required and on time. The distributor is to confirm that these tasks have been executed by 7/31/2023. Potential return of product by the distributors will occur. All end customers within the distributor's territory are to be informed using the Customer Information Letter dated 7/25/2023 (letter #RF-013992-a) which was attached to the Guidance for Sentec Distributors. These letters were to be sent via email using a read receipt. The Customer Information Letter also contained information on the devices impacted, what the defect is, how it manifests itself, and the risks (impact on measurement readings, impact on the sensor membrane, and harm to patient skin). The end user is to identify any lots they may still have at their institution and if any of the affected products are in stock, the recalling firm recommends they follow the visual inspection steps shown in the enclosed Appendix I. If a defective membrane is detected, they are to replace it immediately. The customer is referred to Appendix I and II or the Instruction Manuals listed in the letter for the IFU procedure and picture of what a defective membrane may look like. Enclosed Appendix III contains a table listing the affected lot numbers for the Membran
Initiated
2023-07-25
Root cause
Process control
Distribution
US Nationwide distribution.
Recall number
Z-2592-2023
Recall status
Open, Classified
CDRH record ID
202669
FDA classified
2023-09-15
Recall event number
92833

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗