Skip to content

Public FDA record

Z-2592-2025: Sterile Medline Convenience Kits: 1) MAJOR PACK L-F, Model Number: DYNJ0382730O; 2) LB BASIC CUSTOM PACK

Recall record Z-2592-2025 for Medline Industries, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2592-2025
Company in FDA record
Medline Industries, LP
FDA classification date
Sep 26, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Convenience kits labeled as sterile have not gone through the sterilization process.

Source facts

Action
Customers were alerted via email and telephone beginning 8/8/25 of a potential issue with the products. An URGENT MEDICAL DEVICE RECALL notification letter dated 8/22/25 was sent to customers. This communication is a follow-up to the notification you received from your Medline Sales Representative and to acknowledge the subsequent return of impacted product to Medline. This communication is to provide formal documentation of the recall for your records. No additional action is required to be taken. Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Sterile Medline Convenience Kits. It has been identified that specific convenience kits labeled as sterile have not gone through the sterilization process. As a result, these kits are non-sterile. If the non-sterile kits are assumed sterile upon use, this could result in potential contamination of the sterile-field and an increased risk of patient infection that may require the use of intravenous or oral antibiotics. Refer to the enclosed list for all affected item number(s) and affected lot number(s). Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax and to Medline Industries, LP. If you have any questions regarding this notification, please contact the Director, Regulatory Compliance at (smbrandel@medline.com). or Product Recall Specialist at HBarclay@Medline.com.
Initiated
2025-08-08
Root cause
Process control
Distribution
US Distribution to States of: FL, MD, NJ, TN
Recall number
Z-2592-2025
Recall status
Open, Classified
CDRH record ID
215763
FDA classified
2025-09-26
Recall event number
97495

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗