Open original FDA record ↗Recall record
- FDA identifier
- Z-2594-2011
- Company in FDA record
- Computerized Medical Systems Inc
- FDA classification date
- Jun 20, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Focal: the couch position can be moved between scans of the same CT study. When the images are imported into Focal, the software is not checking the DICOM Image Position (Patient) tag and therefore is not aligning the images sets if there are different couch positions indicated. A shift between image sets will exist of the same amount that the CT couch was moved.
Source facts
- Action
- Elekta CMS Software sent an "ADVISORY NOTICE" dated January 19, 2009 to its affected customers. The notice identified the product, the problem, and the acton to be taken. Customers were informed that the workaround method available is to (use the image fusion feature already in Focal to relign the scans). The firm states that this issue is resolved in Focal version 4.62.00, which was released in March 2011. If you have any questions, please call 314-993-0003 or toll free 800-878-4267.
- Initiated
- 2009-01-19
- Root cause
- Software design
- Terminated
- 2014-04-10
- Distribution
- Worldwide Distribution -- USA (nationwide) including states of: CA, CO, FL, IL, IN, KS, MO, NC, OH, OK, PA, TN, and WA. and countries of Australia, Austria, Canada, India, Netherlands, Portugal, Singapore, Spain, and United Kingdom.
- Recall number
- Z-2594-2011
- Recall status
- Terminated
- CDRH record ID
- 98819
- FDA classified
- 2011-06-20
- Recall event number
- 58253