Z-2595-2018: MAYFIELD SPINE TABLE ADAPTOR RADIOLUCENT, Part Number A2600R The MAYFIELD Infinity XR2 Base Unit is int

Recall record Z-2595-2018 for Integra LifeSciences Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2595-2018
Company in FDA record
Integra LifeSciences Corp.
FDA record creation date
Aug 1, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for the knob assembly in the swivel adapter to fracture/break during use.

Source facts

Action
The firm, Integra, sent an "Urgent Voluntary Medical Device Correction Field Safety Notice" dated 6/29/18 to customers. The notification letter instructed customers to perform the following: 1) Review your MAYFIELD XR2 instructions for use. Please pay particular attention to the CAUTIONS listed on page 5 of the accompanying instructions for use section as noted below. " Always make sure the base unit is properly secured to the operating table " The base unit must not be used if the device appears to be damaged or functioning incorrectly " Over-tightening the mechanisms adjustment screws may result in damage to the unit " Over extending or overloading the base unit may result in unintended movement, shortened product life and/or damage to the unit " Do not alter the construction of this device as it may result in serious patient injury " Do not steam sterilize; components may be damaged by high heat leading to device damage and reduced performance 2) Visit Integra website and review video highlighting instructions for use at the following link: https://www.integralife.com/mayfield-imaging-cranial-stabilization-system/category/cranial-stabilization-mayfield-imaging-cranial-stabilization-system 3) Complete and return the attached Field Safety Notice Acknowledgement Form promptly via email: FCA@integralife.com or Fax to: 1-609-750-4220. Distributors were also instructed to: "We also kindly ask you to contact the final customers who may have the affected products and provide them with this letter." On July 20, 2018, an additional Urgent Voluntary Medical Device Recall letter was distributed to customers. The letters ask customers to perform the following: 1. Allow your sales representative to check the current knob assembly on your MAYFIELD Infinity XR2 Radiolucent Base Unit (A2079), Standard; MAYFIELD Infinity XR2 Low Profile Base Unit (A2079A); MAYFIELD Infinity XR2 Radiolucent Base Unit, Extended (A2079E) and MAYFIELD Spine Table Adaptor (
Initiated
2018-06-29
Root cause
Device Design
Terminated
2020-02-26
Distribution
Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV; and to countries of:: Australia, Canada, China, Dominican Republic, Hong Kong, India, Japan, Korea, Malaysia, Mexico, Pakistan, Panama, Peru, Singapore, Sri Lanka, Taiwan, and Thailand.
Recall number
Z-2595-2018
Recall status
Terminated
CDRH record ID
165399
FDA record created
2018-08-01
Recall event number
80421

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗