Skip to content

Public FDA record

Z-2595-2023: MEDLINE STERILE IRIS SCISSORS CVD/STD, REF DYNJ04049

Recall record Z-2595-2023 for MEDLINE INDUSTRIES, LP - Northfield. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2595-2023
Company in FDA record
MEDLINE INDUSTRIES, LP - Northfield
FDA classification date
Sep 18, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

4.5" Iris Scissors, Sterile, Curved (DYNJ04049) are being recalled due to the potential of the tip protector to fall off. The sharp tip of the scissors could compromise the sterile barrier system by puncturing large and/or small holes that may not be visible to the user.

Source facts

Action
Medline Industries issued a recall notice to its consignees by email on 08/03/2023. The notice explained the problem with the device, the potential hazard, and requested the following: Immediately check your stock for the affected item number and the affected lot numbers listed on the recall portal. Destroy affected product. Distributors or those who resold or transferred the product were directed to notify their customers and have them destroy any affected product. If you have any questions, please contact the Recall Department at 866-359-1704.
Initiated
2023-08-03
Root cause
Under Investigation by firm
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Canada, Hong Kong, Taiwan, Singapore
Recall number
Z-2595-2023
Recall status
Open, Classified
CDRH record ID
202809
FDA classified
2023-09-18
Recall event number
92881

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗