Open original FDA record ↗Recall record
- FDA identifier
- Z-2595-2023
- Company in FDA record
- MEDLINE INDUSTRIES, LP - Northfield
- FDA classification date
- Sep 18, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
4.5" Iris Scissors, Sterile, Curved (DYNJ04049) are being recalled due to the potential of the tip protector to fall off. The sharp tip of the scissors could compromise the sterile barrier system by puncturing large and/or small holes that may not be visible to the user.
Source facts
- Action
- Medline Industries issued a recall notice to its consignees by email on 08/03/2023. The notice explained the problem with the device, the potential hazard, and requested the following: Immediately check your stock for the affected item number and the affected lot numbers listed on the recall portal. Destroy affected product. Distributors or those who resold or transferred the product were directed to notify their customers and have them destroy any affected product. If you have any questions, please contact the Recall Department at 866-359-1704.
- Initiated
- 2023-08-03
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Canada, Hong Kong, Taiwan, Singapore
- Recall number
- Z-2595-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202809
- FDA classified
- 2023-09-18
- Recall event number
- 92881