Skip to content

Public FDA record

Z-2597-2018: power cord for: (a) AMIA Automated PD Cycler, Product Code: 5C9320; (b) Kaguya PD System, Product Code: T

Recall record Z-2597-2018 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2597-2018
Company in FDA record
Baxter Healthcare Corporation
FDA record creation date
Aug 2, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There were instances where the power cord socket detached from the back of the AMIA cycler when the power cord was unplugged from the socket.

Source facts

Action
Method of Notification: A Safety Alert communication will be sent to affected customers via U.S.P.S., first class mail. Mitigating Instructions: Baxter is asking customers to: 1. Unplug the devices from the wall before unplugging the power cord from the back of the cycler. 2. Confirm the receipt of the letter by completing the enclosed customer reply form and returning it to Baxter by either mail, fax, or scanned email.
Initiated
2018-07-12
Root cause
Under Investigation by firm
Terminated
2021-03-31
Distribution
US and Japan
Recall number
Z-2597-2018
Recall status
Terminated
CDRH record ID
165664
FDA record created
2018-08-02
Recall event number
80522

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗