There were instances where the power cord socket detached from the back of the AMIA cycler when the power cord was unplugged from the socket.
Source facts
Action
Method of Notification: A Safety Alert communication will be sent to affected customers via U.S.P.S., first class mail.
Mitigating Instructions:
Baxter is asking customers to:
1. Unplug the devices from the wall before unplugging the power cord from the back of the cycler.
2. Confirm the receipt of the letter by completing the enclosed customer reply form and returning it to Baxter by either mail, fax, or scanned email.
Initiated
2018-07-12
Root cause
Under Investigation by firm
Terminated
2021-03-31
Distribution
US and Japan
Recall number
Z-2597-2018
Recall status
Terminated
CDRH record ID
165664
FDA record created
2018-08-02
Recall event number
80522