The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use.
Source facts
Action
The firm, BD Life Sciences, produced the "URGENT PRODUCT RECALL-CORRECTION" letter dated 4/6/2018 to its customers during an on-site visit. by the field service engineer made beginning 5/16/2018 to correct the issue. The visits were completed 6/15/2018.
If you require further assistance, contact BD Customer/Technical Support
BD Customer Support at 877.232.8995 (prompt 2) for instruments in the United States. For customers outside the US, contact your local BD
representative or distributor.
Initiated
2018-05-16
Root cause
Nonconforming Material/Component
Terminated
2018-08-13
Distribution
US Distribution was made to FL, KS, KY, MA, NJ, and SD.
Recall number
Z-2598-2018
Recall status
Terminated
CDRH record ID
165331
FDA classified
2018-07-13
Recall event number
80388