Skip to content

Public FDA record

Z-2599-2018: MicroScan Pos Combo Panel Type 43, Catalog number B1017-217for use in determining antimicrobial agent sus

Recall record Z-2599-2018 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2599-2018
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
Aug 2, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The drug dilution sequence for Moxifloxacin printed on the panel is incorrect. As a result, incorrect MIC results for Mxf could be reported.

Source facts

Action
A recall notification letter was sent the week of 05/01/2018 via email, mail, and phone if needed.
Initiated
2018-05-04
Root cause
Labeling design
Terminated
2020-08-07
Distribution
NJ, TX, IA, PA, FL, NE, OH, NY, CA, MO, OK, LA
Recall number
Z-2599-2018
Recall status
Terminated
CDRH record ID
165211
FDA record created
2018-08-02
Recall event number
80330

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗