Skip to content

Public FDA record

Z-2600-2011: Prism Laser 7642 Laser System

Recall record Z-2600-2011 for American Stress Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2600-2011
Company in FDA record
American Stress Technologies, Inc.
FDA classification date
Aug 29, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

It was discovered The Prism 7642 was not certified at the time of delivery. The product lacks the required labeling and the product was sold and delivered without the submission of the Initial Product Report. The product lacks User In formation

Source facts

Action
American Stress Technologies sent a customer notification letter dated April 1, 2011, informing them of the action they have taken to correct this issue. (1) ASTI conducted laser safety tests and checks on May 3,2011 to verify the Prism was compliant with FDA regulations. This was confirmed in a June 17, 2011, electronic mail message. (2) ASTI placed the required labels described in the initial product report on the Prism on May 3, 2011. (3) ASTI delivered updated user manuals to UNO on May 3, 2011, that describe potential laser hazards and the required labeling in the same way as described in the initial product report. (4) ASTI completed the above corrective actions at no cost to UNO as confirmed in a June 20, 2011, electronic mail message. If you have any further questions please call (724) 410-1030
Initiated
2011-04-27
Root cause
Radiation Control for Health and Safety Act
Terminated
2016-10-13
Distribution
Nationwide Distribution
Recall number
Z-2600-2011
Recall status
Terminated
CDRH record ID
100331
FDA classified
2011-08-29
Recall event number
58804

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗