Open original FDA record ↗Recall record
- FDA identifier
- Z-2600-2018
- Company in FDA record
- Encore Medical, Lp
- FDA record creation date
- Aug 2, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A black, aluminum oxide coating may appear on the tray after cleaning. The black, aluminum oxide coating is not biocompatible.
Source facts
- Action
- The firm, DJO LLC, sent an "URGENT FIELD SAFETY NOTICE" letter dated 6/26/2018 to its direct accounts by email on 06/26/2018. The letter described the product, problem and actions to be taken. The customers were instructed to return product; contact Customer Service as soon as possible to request an RMA for return and exchange of the full instrument set; and to coordinate the exchange of products with Customer Service. If you do not have these instrument sets or the number of sets identified above, you are required to complete the attached form and email it to sandra.busick@djdoglobal.com. ALL ACTIVITIES OUTLINED ABOVE SHOULD BE COMPLETED BY JULY 30, 2018. If you have any questions, please call (512) 834-6255 or by email to teffany.hutto@djoglobal.com.
- Initiated
- 2018-06-26
- Root cause
- Process control
- Terminated
- 2023-08-22
- Distribution
- US Distribution to states of; AZ, CA, CO, FL, IN, NJ, NY, NV, MI, MN, MO, OH, RI, SC, TN, TX, UT and WA and country of; Japan.
- Recall number
- Z-2600-2018
- Recall status
- Terminated
- CDRH record ID
- 165508
- FDA record created
- 2018-08-02
- Recall event number
- 80445