Open original FDA record ↗Recall record
- FDA identifier
- Z-2601-2016
- Company in FDA record
- Siemens Healthcare Diagnostics Inc
- FDA classification date
- Aug 19, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Some v2.2.2 upgrade kits include a dialysate mode which not cleared/approved for shipment in the United States
Source facts
- Action
- Siemens sent a Customer Notification dated July 2016 to all affected domestic (US) customers via Federal Express on July 5, 2016. Customers were asked to take the following actions: Please do not use the dialysate mode. If you have enabled this feature, please disable it by following these instructions: 1. At the System screen, select Setup. 2. Select Secured Options > Analysis Options. 3. Unselect Dialysate. 4. Select the Continue button 3 times. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check attached to this letter within 7 days. Customers with questions were instructed to contact their Siemens Customer Care Center or their local Siemens Technical Support representative. For questions regarding this recall call 781-269-3000.
- Initiated
- 2016-07-05
- Root cause
- Under Investigation by firm
- Terminated
- 2017-08-02
- Distribution
- Nationwide Distribution to AL, CA, MA, NJ, OH, TX, WA, PA
- Recall number
- Z-2601-2016
- Recall status
- Terminated
- CDRH record ID
- 147912
- FDA classified
- 2016-08-19
- Recall event number
- 74564