Z-2601-2021: LikoStretch Mod 600 IC Model Number: 3156065 LikoStretch Mod 600 IC, Wide Model Number: 3156065B (obsolet

Recall record Z-2601-2021 for Hill-Rom, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2601-2021
Company in FDA record
Hill-Rom, Inc.
FDA record creation date
Sep 30, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from.

Source facts

Action
Hillrom issued URGENT Medical Device Correction letter dated August 11, 2021 by third party vendor via UPS mail/track the delivery to consignees. Letter states reason for recall, health risk and action to take: Hillrom is initiating this Medical Device Correction to all customers and will be correcting devices with an updated to the leveller. Actions being taken by Customer: Please share this Medical Device Correction with all potential users in your organization. Please ensure inspection of the device prior to use as stated in the device Instruction for Use (IFU). Complete the attached response form, acknowledging receipt, indicate the number of devices and return to hillromLUL003@stericycle.com within one month
Initiated
2021-08-11
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DE, IL, ME, MI, MO, MS, NJ, OH, PA, VA, WA and the countries of Argentina, Australia, Austria, Bolivia, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Latvia, Lithuania, Mexico, Netherlands, Norway, Poland, Portugal, Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Turkey, United Arab Emirates.
Recall number
Z-2601-2021
Recall status
Open, Classified
CDRH record ID
189067
FDA record created
2021-09-30
Recall event number
88548

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗