Skip to content

Public FDA record

Z-2603-2016: TIGR Matrix Surgical Mesh, Model No. NSTM1015, NSTM1520, NSTM2030

Recall record Z-2603-2016 for Novus Scientific Ab. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2603-2016
Company in FDA record
Novus Scientific Ab
FDA classification date
Aug 19, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Novus Scientific is implementing additions to the product's instructions for use concerning certain inguinal hernia repairs.

Source facts

Action
An Urgent Safety Alert letter dated 7/19/16 was sent to all their customers to inform them that Novus Scientific is implementing additions to the product's instructions for use (IFU) with a new contraindication and warning related to the clinical study result. The letter informs the customers that until the new IFU is made available, surgeons using the product for hernia repair should follow the instructions in the urgent safety alert in addition to those in the current IFU. Customers are informed of the actions to be taken: " Additional contraindication: Not suitable for the repair of direct inguinal hernias. " Additional warning: Because TIGR Matrix Surgical Mesh is fully resorbable, it should not be used in repairs where permanent support from the mesh is required. Customers with any questions or concerns are instructed to contact Novus Scientific AB ( +46 (0)18 700 11 50.
Initiated
2016-07-19
Root cause
Other
Terminated
2016-12-23
Distribution
US : IL, CA, FL, KY, MO, NH, VA, TX, NC, OH, NY, TN, Worldwide: Sweden, Italy, Denmark, UK, Singapore, Germany, Belgium, Spain, Austria, Israel, Malta, Switzerland, Ireland, Norway, Netherlands, Turkey, New Zealand, Bulgaria, France, Iran, Finland, Poland, South Africa
Recall number
Z-2603-2016
Recall status
Terminated
CDRH record ID
148479
FDA classified
2016-08-19
Recall event number
74801

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗